Because of a lapse in government funding, the information on this website may not be up to date, transactions submitted via the website may not be processed, and the agency may not be able to respond to inquiries until appropriations are enacted. The NIH Clinical Center (the research hospital of NIH) is open. For more details about its operating status, please visit cc.nih.gov. Updates regarding government operating status and resumption of normal operations can be found at opm.gov.

ORF Home > Technical Resources > Policies and Rules > Fume Hood Specifications

Fume Hood Specifications


The work required under this section consists of providing and installing laboratory hoods with horizontal sashes and related items necessary to complete the work as indicated on the drawings and as described herein. Hoods shall be bench type and 1800 mm (nominal 6 ft.) wide as shown.

15991 - Construction contract specifications for constant volume fume hood testing PDF file
 - PDF (235KB) 508
Tests identified below, other than the ANSI/ASRAE 110 test, were created by Farhad Memarzadeh of the National Institutes of Health in 1997 and further revised by Memarzadeh and Brightbill in 1999.   On site testing shall not start until testing, adjusting and balancing of the air and water systems, calibration and tuning of controls systems, off site testing of fume hoods and commissioning are complete and the facility is ready for occupancy.

11820 - Construction contract specifications for horizontal sash fume hoods PDF file - PDF (115KB) 508
The work required under this section consists of providing and installing laboratory hoods with horizontal sashes and related items necessary to complete the work as indicated on the drawings and as described herein. Hoods shall be bench type and 1800 mm wide (nominal 6-foot) as shown.

15992 - Construction contract specifications for variable volume fume hood testing PDF file - PDF (293KB) 508
Test identified below, other than the ANSI/ASHRAE 110 test, were created by Farhad Memarzadeh of the National Institutes of Health in 1997 and further revised by Memarzadeh and Brightbill in 1999.  On site testing shall not start until testing, adjusting and balancing of the air and water systems, calibration and tuning of controls systems, off site testing of fume hoods and commissioning are complete and the facility is ready for occupancy.

11810 - Construction contract specifications for vertical sash fume hoods PDF file - PDF (253KB) 508 
The work required under this section consists of providing and installing bypass laboratory hoods with vertical sashes and related items necessary to complete the work as indicated on the drawings and as described herein. Hoods shall be bench type and either 1200 mm or 1800 mm (nominal 4-foot or 6-foot) as shown on drawings.

 